Log entries with a label of licor.

32m sonic was down
32m statistics were down (sonic, licor, pressure) at 32m from about 0400-0700 this morning.  They are back now, either by themselves or by some action from Gordon. This is related to a second rain event last night.
cleanings
Well....I <almost> made it.  (My first time high on the tower.) Was on the tower from 1055-1210 today, cleaning everything from 2m->32m.  As Chris did, I started by putting a drop of water on the wetness sensor (with the exception of the 2m level, which doesn't have one, and the 8m licor that I did just a minute before putting the drop on).  Later, I realized that I could use the drop as a source of water while doing the cleaning!</almost>…
20m Licor Fixed
Found the problem with the 20m licor was the serial cable from electronics to dsm.  Bulgin connector at dsm was full of water.  Ran a new serial line from Licor to dsm.  All is good!
Loaner licor (0287) is up and reporting.
32m Licor Removed
Removed the 32m Licor.  Waiting on replaced (loaner) from Licor and shipping bad back to company.
Licor 7500 at 32m
Licor at 32m stopped reporting again this morning. Power cycling didn't help: ssh mid eio 9 0 eio 9 1 Also tried minicom, sending a BREAK (ctrl-A F) and the Outputs command. No response. Here's the LICOR 7500 configuration for SOASl, output at 20 Hz, bandwidth 20Hz: (Outputs (BW 20) (RS232 (Freq 20.…
Licor at 32 m is flaky
June 5: 11:28 - 11:49 The Licor at 32 m (SN 1164) has been intermittent since the heavy rain on the afternoon of June 1.  Gordon found the "ready when off" light to be steady red.  The supply voltage was only 8.04 V, while the same voltage for the Licor at 38 m read 10.8V. Chris suggested swapping heads to further diagnose the problem.
Tower sensor maintenance
June 5: ~ 11:25 - 12:40 Visited tower to service 32 m Licor and 8m mote.  Also added more bird-b-gone to top of tower.
32m Licor restarted
June 3: h2o.32m (licor) appears flaky starting around 17:00, June 2. Around 12:00 CDT, June 3, Gordon restarted the 32 m Licor over an rserial connection.
Tower service visit
May 30, ~ 10:30 - 1:30 CDT Gordon and Tom serviced the sensors and DSMs on the profile tower: 44m: Cleaned Licor 38m: Cleaned EC150 ~36m: Cleaned K&Z 4-component net radiometer 32m: Cleaned Licor; grounded DSM; cleaned PAR; Tsfc @ -1.5 degC 26m: Cleaned Licor, PAR, and wetness sensor (splattered w/ bird crap) 20m: Licor and PAR cleaned; PAR covered in bird crap 14m: Grounded DSM and fixed usbdisk; cleaned PAR, KH2O (0.35V) and wetness sensor; no Tsfc 8m: Cleaned PAR, Licor,…
45m Turbulence Tower Height Sonic TRH Scalar Radiation Inlet Other 44m 1122 030(022) Li1163 IRT3389-1   Pnano-123997 38m 1244 026(029) EC1313 CNR4-090020, Lwet     32m 0672 009(065) 1164 PAR 6951, IRT3389-3, Lwet   Pnano-123996 26m 0673 007(053) 1167 PAR1892, IRT3577-2, Lwet     20m 1121 29(051) 1166 PAR1893, IRT3389-1, Lwet   Pnano-123998 14m 0671 (0800) 40(030) krypton 1101 PAR1529, IRT4078-3, Lwet     8m 0800 (0538) 60(041) 0813 PAR1530, IRT3577-4,…