Andy INTERCOMPARISON SYSTEM FIELD NOTES, --------- Friday, September 24, 2004 ------- (CU and HYDRA Times in watch time MST, USGS Time in Datalogger time, AIRCOA A and AIRCOA B times are Prometheus time ) AIRCOA B Aspen site 15:56 UTC Currently doing leak check 16:00:12 Took Control. Turn on Guest Cal. Cal A on, V12 on 16:00:45 P = 71.77, F = 1.41, CO2 = 395.0 - 395.4 16:00:51 V12 off, slow flow, P = 71.77, Flow = 1.41, CO2 = 391.1 - 392.5 16:03:05 Switch to B, V12 on. 16:03:23 P = 80.061, F = 3.00, CO2 = 324.3 - 324.9 16:03:56 V12 off. 16:04:30 P = 71.86, F = 1.43, CO2 = 320.9 - 321.3 16:06:11 Switch to C, V12 on 16:06:45 P = 79.79, F = 3.02, CO2 = 355.1 - 356.3 16:06:50 V12 turned off. 16:09:00 P = 71.82, F = 1.53, CO2 = 350.2 - 351.3 16:09:20 "r" to resume program Cal cylinder pressures: HS2 = 1550, HS1 = 1590, LS1 = 1700, LS2 = 1490. CU Tower 10:58:00 MDT Currently on Line 3. Reading is 1.323 10:58:40 Switched profile to manual. Toggled C2. P = 20.5 psig. 11:01:30 3-way to ICS, Cylinder A 11:02:50 Prssure good at 20.5 psig 11:06:00 Step 4-way to B 11:07:00 Pressure good at 20.5 psig, Vco2 = 1.140 11:10:18 Step 4-way to C 11:11:00 Pressure good at 20.5 psig, Vco2 = 1.212 11:15:20 Switched profile to auto. Toggled C2. USGS Tower 10:26:59 MST Datalogger time; watch time is 11:28:00 10:27:30 Switch sampling from 11 to 12, P = 547.8 - 531.5 10:29:30 Switched 3-way to start ICS, Cylinder A 10:30:30 Pressure good at 537.1, CO2 = 392.7 10:32:30 Step 3-way to B 10:33:00 Done adjusting regulator to get 545.1, CO2 = 324.9 10:35:00 Step 3-way to C. 10:35:30 Done adjusting regulator to get 543.5, CO2 = 353.7 10:37:30 Switched 3-way back to sample from towers, line disconnected. HYDRA at Pine Tower Connect to 3-way valve and quick connect between HYDRA and Instrument box On line Sx-S2. Sample Ball Flow = 36. Pump Pressure = 8.4 psi 12:07:14 Switch to gas A 12:07:45 Done adjusting regulator to get Flow = 36, Pressure = 8.4 psi CO2 = 61.4 um/m 12:10:00 Step 3-way to B 12:10:15 Adjusted regulator to get Flow = 36, Pressure = 8.4 psi CO2 = 61.4 um/m 12:12:30 Step 3-way to C 12:13:00 Adjusted regulator to Flow = 36, Pressure = 8.4 psi CO2 = 36.5 um/m 12:15:30 3-way back to HYDRA, System on D-S4 AIRCOA A Willow site 2 m line selected, Slow flow = 1.616 - 1.688, Fast flow = 3.07 - 3.09 19:43:50 Turn on Cal C, fast flow = 3.04 - 3.07 19:44:35 V12 off 19:45:23 P = 72.265, F = 1.622 - 1.699, CO2 = 327.7 - 326.62 19:47:13 Step 3-way to B, fast flow 19:47:45 P = 80.404, F = 3.05 - 3.07, CO2 = 303.14 - 304.43 19:48:56 Slow flow 19:49:47 P = 72.19, F = 1.61 - 1.83, CO2 = 299.18 - 300.24 19:51:13 Step 3-way to A, fast flow 19:51:32 P = 80.363, F = 3.01 - 3.02, CO2 = 369.29 - 369.84 19:51:45 Slow flow 19:52:52 P = 72.194, F = 1.61 - 1.68, CO2 = 365.52 - 366.38 19:54:07 'r' to return control to Prometheus. Cal cylinder pressures: HS2 = 1580, HS1 = 850, LS1 = 1660, LS2 = 1730 AIRCOA C Willow site Fast flwo pressure = 71.599 22:10:28 Turn on Cal A, V11, V12 on 22:11:59 Prssure good 22:12:30 V12 off 22:13:09 P = 71.532, F = 1.64, CO2 = 397.9 - 400.7 22:14:18 CO2 = 390.6 - 395.1 22:14:54 Step 3-way to B, V12 on 22:17:23 P = 78.877 22:17:48 V12 off 22:18:42 P = 71.573, F = 1.72, CO2 = 318.9 - 321.6 22:19:39 Step 3-way to C, V12 on 22:21:40 Pressure good at 78.662 22:22:14 Fliped V12 a bit early, so toggled it for a little more flush flow, then off 22:23:23 P = 71.532, F = 1.67, CO2 = 347.0 - 350.8 22:24:45 'r' to return control to Prometheus