board

SN

corrected clock rate (Hz)

0

302859

20,000,292.73

1

302828

20,000,450.67

2

302832

20,000,284.66

3

302871

20,000,072.66

4

303835 302835

20,000,242.86

Pressure2

board

SN

corrected clock rate  (Hz)

0

302821 ?

20,000,314.13

1

302829 ?

20,000,364.39

2

302826 ?

20,000,354.37

3

302874 ?

20,000,342.21

4

302842 ?

20,000,346.47

GPIO-MM Clock Calibrations

...

The reference frequency was stepped in 10 KHz increments from 110 to 300 KHz approximately every 10 seconds. The GPIO cards were configured to count the number of  tics from the onboard 20 MHz clock while counting 10,000 cycles of the reference input (, which is the same manner that the frequencies of the ParoScientific temperature transducers are sampled). These counts are repeated every 0.1 seconds.  The resulting frequency measurements were then compared against the reference frequencies.

...