Post cals were conducted on the SHT probes. Temperature results were within +/- 0.1C. All sensors were within .05C except for sensors SHT001 (error: -.08C) and SHT006 (error: .07C). The humidity is a different story. Over the range of 10% to 90% the average error for all the sensors went from 1% to -8% (figure 1).