Per Steve Oncley, increased the sampling rate of the Licor 7500 at 43 meters.
It is the only unit connected right now.

Commands to change the sampling rate:

# shut down data process
adn

# adn powers off port 17, power it back on.
eio 17 1

# start minicom
# The default baud rate of minicom is 9600, which is what the 7500 uses after it is sent a break.
# To set it to 9600 explicitly, use ctrl-A pe.
# Use ctrl-A sequence to turn on local echo (E), line wrap (W) and send a break(F).
# Set bandwidth to 20 and RS232 output frequency to 20:

minicom ttyS17
ctrl-A pe
ctrl-A e
ctrl-A w
ctrl-A f
(Outputs BW(20) (RS232 (Freq 20.0)))
(Outputs ?)
ctrl-A q

# minicom should now be finished

# power off port 17
eio 17 0

# start data process. This will power up port 17
aup

# wait a bit, then
ds
ctrl-C


# If the 7500 doesn't respond, try power cycling it:
eio 17 0
eio 17 1