RF05
5/27/25
Tech: Ivana
Pre-flight
-Scattered showers during pre-flight (radome tape and probe covers on until APU start)
-VSCEL showing NaN value after 2 reboots, likely due to rain, so will turn on after takeoff
Flight
-0152z: VSCEL ON and operating properly
-DPL dewpoint value and 2DC diode voltages look nominal
-Left OH imager missing some data (~500 images), likely due to errant VNC clicking. Catherine will look into making VNC on L-6 laptop view-only
-0825z: Re-filled dry ice at R-5 RackĀ