• No labels

3 Comments

  1. Cleaned visible and Heimann radiometers per the request of Julie Haggerty.

    Janine put in fixes for the winds system.  Also, enabled DGPS from the Novatel for Stuart Beaton.

    After full data system checkout, QCF was not reporting properly.  Symptom was red samples/sec on status page.  An occasional 20 samples/sec would pop up but didn't match any of the Honeywell's on that serial card.  Power cycled 5 times with no success.  rserial into it's serial channel and looked like garbage data.  Ran serial loopback test on all serial channels on that card and all tested good.  Swapped with spare Honeywell and all looks good.  It may have reverted back to an old baud rate.  Will need to be checked out back at RAF.  If there is another spare, possibly the next tech could bring it down with him.  Original S/N 33821 and was swapped out with the spare S/N 47172.

    Decided to abandon a CDP cal due to intermittent rain.  Will attempt next maintenance day.
     

    A2D auto calibration had to be cancelled due to lightening nearby.  Decided to stop for the day.  We are ready for RF02.

  2. Kyle,

    And or maybe the honeywell reverted back to it's old initial setting, similar idea.  Glad you got it resolve!

    Tks

    1. Unknown User (munnerly)

      Pressure transducer S/N 33821 evaluated in lab using a computer with Hyperterminal and a lab supply. Transducer has some functionality with its factory settings of 9600, 8N1, none. It does not however, respond to a pressure query (*99P1=) at any baud rate. Will enter as "broke beyond repair" in database, harvest connectors and remove from service.