Checked radome for moisture in pressure lines. Moisture present in top ADIFR line only. Blew out top ADIFR line. Opened all lines to let them air dry - just because.
Performed CDP Cal
Justin pulled the PCASP probe to replace the dessicant
Justin pulled the CN plate to work on the pump flow meter
-Inverters remained on and functional throughout duration of maintenance day
-Installed AMS Computer in R3 Outboard Rack. Issue with AMS Top UPS. Battery connected outputs not providing power. Display / Primary Power Switch is not working. Manual indicates bad battery. Solution is to bypass the upper UPS by connecting TOFS Power Supply and Computer 60 Hz cables to power strip on 60Hz circuit 2.
-Camera still not sending images to the ground; Janine has contacted Chris and Carol since it is an LDM issue that came up during power drop yesterday
-L-10 Ozone instrument working properly again after having issues yesterday after power drop
-Inspected radome lines - no moisture
-Inspected VSCEL - cleaning not needed
-Changed CO2 calibration cylinder (Right #1)
-L-10 Inboard Aeros re-installed and operational (old Aeros configuration files saved to Desktop > Aeros config)
13 Comments
David Allbee
Jul 18, 2025Performed CDP Cals
Cleaned CDP windows
Cleaned 2D-C mirrors
Cleaned lenses on RSTB, RSTB1, and RSTT
Cleaned IRT and VIST domes
Cleaned VCSEL
Flushed Temp Probes with water
UHSAS ethernet cable repaired - tab needed to be bent upward, now it seats into the network hub properly
RAF CN plate removed and opened up by Justin for cleaning. Critical orifice was cleaned and put back together
RAF CN filled with butanol
New exhaust line ran from the ISAF instrument to window exhaust port on right side bulkhead
New instrument (temp controller) installed in R-6 rack
David Allbee
Jul 21, 202507/19/25
Power and Access / User Support
David Allbee
Jul 21, 20257/21/25
Pulled PCASP Probe for Justin. He changed the desiccant in the drier tube. Then reinstalled and tested.
David Allbee
Jul 27, 20257/26/25
Cleaned RSTB, RSTB1 and RSTT
Cleaned IRT and VIST domes
Cleaned 2D-C mirrors - residue removed from both
Cleaned VCSEL mirrors - a tiny pit is visible in the lower mirror
Flushed both temp probes with water
Cleaned DPT and DPB
Checked radome for moisture in pressure lines. Moisture present in top ADIFR line only. Blew out top ADIFR line.
Filled RAF CN with butanol
David Allbee
Jul 28, 20257/28/25
Performed CDP Cals
Cleaned CDP windows
Power and Access / User Support
David Allbee
Jul 31, 20257/31/2025
Power and Access / User Support
David Allbee
Aug 01, 20258/1/2025
Cleaned CDP windows
Cleaned 2D-C mirrors
Cleaned RSTB, RSTB1 and RSTT
Cleaned IRT and VIST domes
Power and Access / User Support
David Allbee
Aug 07, 20258/7/2025
Cleaned RSTB, RSTB1, and RSTT
Cleaned IRT and VIST domes
Cleaned CDP windows
Cleaned 2D-C mirrors
Cleaned VCSEL mirrors
Flushed both temp probes with water
Cleaned DPT and DPB
Checked radome for moisture in pressure lines. Moisture present in top ADIFR line only. Blew out top ADIFR line. Opened all lines to let them air dry - just because.
Performed CDP Cal
Justin pulled the PCASP probe to replace the dessicant
Justin pulled the CN plate to work on the pump flow meter
Ivana Vu
Aug 29, 20258/13/25 Maintenance
-Power and access for users
-Replaced PCASP desiccant beads
-Cleaned bottom camera lens
-Cleaned RSTB & RSTB1
-Cleaned CDP lens
-Cleaned VECSEL (laser intensity measured at 1250 after cleaning)
-Removed CV #3 (4% SO2)
-2DC Diode Voltages: DIOV01=3.13 , DIOV32=3.5 , DIOV64=2.88
-Blew out DPT and and DPB
-Tested ICS Station #11 with new push to talk switch. No issues found with audio levels when speaking.
Ivana Vu
Aug 29, 20258/17/25 Maintenance
-Power and access for users
-Replaced PCASP desiccant beads
-Cleaned RSTB, RSTB1, RSTT, Top Stabilized Platform Radiometers
-CDP Calibration
-Changed CV #3 (4% SO2) and CV #4 (0.5% SO2)
-Cleaned DPB (dewpoint flag and head)
-Changed TOGA Helium cylinder
Ivana Vu
Aug 29, 20258/19/25 Maintenance
-Power and access for users
-Inverters remained on and functional throughout duration of maintenance day
-Installed AMS Computer in R3 Outboard Rack. Issue with AMS Top UPS. Battery connected outputs not providing power. Display / Primary Power Switch is not working. Manual indicates bad battery. Solution is to bypass the upper UPS by connecting TOFS Power Supply and Computer 60 Hz cables to power strip on 60Hz circuit 2.
-Camera still not sending images to the ground; Janine has contacted Chris and Carol since it is an LDM issue that came up during power drop yesterday
-L-10 Ozone instrument working properly again after having issues yesterday after power drop
-Inspected radome lines - no moisture
-Inspected VSCEL - cleaning not needed
-Changed CO2 calibration cylinder (Right #1)
-L-10 Inboard Aeros re-installed and operational (old Aeros configuration files saved to Desktop > Aeros config)
-Changed PCASP desiccant beads
-Removed and re-installed HARP for calibration
Ivana Vu
Aug 27, 20258/26/25 Maintenance
-Power and access for users
-Cleaned VCSEL, RSTB, RSTB1, RSTT, and SP Radiometers
-Removed and reinstalled CV #3 for HOX
-Removed and reinstalled CN Counter plate for scientist cleaning in the lab (cleaned critical orifice)
-Changed PCASP desiccant
-Changed Zero Air cylinder for CO2 instrument (Right #3)
Ivana Vu
Aug 29, 20258/29/25 Packing Day
-Power and access for users
-Packed containers and prepared for shipment (to be picked up on Tuesday 9/2 and arrives at RAF on Friday 9/5)
-Removed R-6 A-ATOFMS cabling
-Removed R-1 TRAC instrument