8/23/25
Tech: Ivana

Takeoff 0740
Landing 1305

Pre-flight
-Changed N2 and O2 cylinders for HOXCIMS (Left #1 & 4)
-Re-installed CV #1, CV #2, and CV #3
-LDM re-started so ground should be receiving camera images now

SATCOM 
0742 Dropped
0801 Reset
0808 Recovered
0854 Dropped
0900 Reset
0910 Recovered
0957 Dropped
1000 Reset
1007 Recovered
1051 Dropped
1053 Reset
1103 Recovered
1146 Dropped
1153 Reset
1158 Recovered
1245 Dropped
1249 Reset


Flight
0934 Noticed that Nimbus terminal window was blank (not showing IWG feed)
0938 Accidentally closed data acquisition window while trying to copy text inside terminal with Cntl+C; Immediately restarted data acquisition ("continue flight")  and Nimbus; Users did not report data gap or issues with Aeros in the past hour except for TOGA; TOGA was revived after ~10 minutes but is unsure if instrument issue was correlated with restarting data acquisition; Confirmed that ground is still receiving data feed

Post-flight
-All data (before and after nimbus was restarted) uploaded properly and was automatically pieced together using copy_data script on ADS desktop
-Janine confirmed that data from RF18 has ~10 second data gap
-Next time, if Nimbus screen is not showing IWG feed, go to Controls Tab > Status > Nimbus to confirm it's green