Dan had mentioned last night that the barometers might still be in 7E1 mode and thus not correctly talking to the DSMs.  This was correct for s10.  I have just logged in, turned off data acquisition so that I could use minicom to change the serial port configuration, and then was able to issue the .8N1 and .RESET commands.  I then started data with a .BP.  I did restart data acquisition (dup).

Since s4 wasn't connecting, I haven't been able to issue the same commands yet to it.


2 Comments

  1. Steven Oncley AUTHOR

    Just did the same for s15.