To setup the Licors for the format that our data system expects, do the following from the data system login prompt:

  1. shut down the data process
    adn
  2. run minicom on the Licor port (ttyS2=2m, ttyS7=7m, ttyS11=16m, ttyS14=30m, ttyS17=43m)
    minicom ttyS2
  3. If the baud rate agrees with minicom (9600) you should see the default Licor output
  4. Send a break character, by doing "control-A f" from minicom
    ctrl-Af
  5. Send these strings to the Licor:
    Code Block
    (Outputs (RS232 (EOL "0A") (Labels FALSE) (DiagRec FALSE) (Ndx FALSE) (Aux FALSE) (Cooler FALSE)
    (CO2Raw TRUE) (CO2D TRUE) (H2ORaw TRUE) (H2OD TRUE) (Temp  TRUE) (Pres TRUE) (DiagVal TRUE)))
    (Outputs (BW 10) (RS232 (Freq 10.0)  (Baud 9600)))
    
  6. You won't see much output, perhaps just 1 column of digits on the right-hand-side, since the Licor is sending just a line-feed termination and no carriage-return. You can turn on line wrap in minicom, with "control-A w" to see the data stream wrap from side to side:
    ctrl-Aw
  7. Exit minicom with control-A q
    ctrl-Aq
  8. restart the data process:
    aup
  9. check the Licor data with rserial. It may take 30 seconds or so for the data system to open the port. During that time rserial will report an error about not finding the sensor. Eventually it should work.
    rs 2
  10. The data should look like:
    Code Block
    248\t0.06460\t12.4193\t0.01218\t81.586\t16.10\t76.1\t\n
    248\t0.06462\t12.4179\t0.01207\t81.459\t16.11\t76.1\t\n
    
  11. Exit rserial with control-D
    ctrl-d