Radiometers: not getting rpile.in, tcase.in, others reporting (tried pushing scan i2c button, didn't fix) → swapped in radiometer with right id for longwave in, now all reporting.

NR01: updated PIC code and plugged into uw.  Isabel updated the configuration so data are expected.

Gsoil/qsoil/lambda: not yet plugged in (Steve is updating PIC code on these sensors)with latest PIC code. Getting data from all, qsoil values are slightly negative. Is this a problem?

Tsoils: all reporting, data looks good.

...

TRH: all reporting (just replaced fuse + housing for 13m trh, it's reporting now but not yet making it to qctables), data looks good.

CSAT: no winds from c 5, 10, 15, 20m. Haven't done much looking into this yet. Looked at c 15m for a few minutes, lights in EC100 box are all green but not getting any data over the port (cu port 7). Will keep looking into the others.

  • ue3m - sonic + gas lights are red. getting messages from sonic, but uvwtc, co2/h20, tirga/pirga are all nans. → after power cycle w/ pio lights are green + reporting winds + gas
  • ue10m - same behavior as ue3m, fixed same way
  • c5m - lights are green but no data (at all) on port. power cycle didn't help. note on ec100 that the pelican case it came out of had been labeled 'needs output reformat'
  • c10m - red for sonic, green for gas and power. gas data ok, nans for winds (but then why is gas data also missing on qctables?). sonic diagnostic is 16, or 'acquiring' → took sonic out of box to see if something was in the way, but no improvement.
  • c15m - see above
  • c20m - sonic light flashing red/green, others green. not sending any data to dsm.

EC150: no gas from all the ones listed above, plus uw 10m. All the co2 values from the EC150s that are reporting are clipped in qctables.

Pillows: currently no data. Did these get unplugged? I know we were getting data from them earlier...

all reporting after I swapped the sonic head on 10m.c (and previous resetting of some EC100's using ECmon).  Some heads are intermittent, but I presume this will clear up in the field.

EC150:

  • all h2o reporting.  Some heads were blocked by foam padding.  (3m.uw died over friday night)
  • alll co2 reporting, but about half are clipped.  Hoping this will go away in the field.
  • Note that, in the ECmon reconfiguring, I noticed 2 EC100 boxes that were set to force Tirga to 25C.  I think these are all fixed now.

Pillows: working, needed reboot of cl. There are only 2 boxes connected.   Saturday: Now back down to 1 pillow; pillow 2 doesn't respond, even manually.  Don't know why not.

Flowcaps: getting data from both flowcaps.  The second needed to have the power switch turned on its USB adaptor.  Both are now configured to the averaging we want (30/30/300)Flowcaps: getting data from 1m flowcaps, not from 2m. Do we have a second flowcaps to stage, or is it just the 1m one? Probably just the one.

Apogee IR: all reporting. Raw values look good, Steve is working on debugging the surface temp calculation.

Tsnows: The one tsnow currently Currently plugged in looks good. Are we planning to stage any of the others? Yes, once Chris updates the sensor ID numbers.to d and ue.  (uw is done, but Chris is protecting the sensors so the cable doesn't reach.)  Looks good. 

Gps/chrony: finally started getting good GPS on uw, but just recently stopped getting good gps on ct. I'll look into it, see jira for more details. → ct fixed itself after I pulled the GPS antenna out to rest on the top of the dsm box.

Victrons: not staged, not planning to stage (actually want to stage at least one as part of a power drop test – Tony is building up power boxes for this)Ott: not yet staged, are we staging this?

Anything else I'm forgetting??