Following the SCP project, we tested each sonic in a zero-wind enclosure within the EOL temperature chamber, measuring wind component offsets and sonic temperature errors over the nominal range -30 C to 55 C.  The following table shows wind component offsets as okay if the u and v offsets are less than +/- 4 cm/s and the w offsets less than +/- 2 cm/s (between - 20 C and 30 C).  The sonic temperature corrections are listed as the slope and offset of a linear fit (between -20 C and 30 C), Tc = offset + slope*tc.sonic, where Tc is calculated from temperature, relative humidity, and pressure measured in the zero-wind enclosure.

Sonic

Location

Cal date

u

v

w

tc offset
(degC)

tc slope

0923

Ah1

 

 

 

 

 

 

0671 cg
0833 gm

Ah2

 

 

 

 

 

 

0743

Aph3

 

 

 

 

 

 

1120

A4

 

 

 

 

 

 

0732

Ah5

 

 

 

 

 

 

0800

Ah6

 

 

 

 

 

 

0673

A7

 

 

 

 

 

 

0176

Ap8

 

 

 

 

 

 

1121

Ap9

 

 

 

 

 

 

0677

Ap10

 

 

 

 

 

 

0674

Ars11

 

 

 

 

 

 

0855

Ap12

 

 

 

 

 

 

0745

A13

 

 

 

 

 

 

1124

Ap14

 

 

 

 

 

 

1122

A15

 

 

 

 

 

 

0740

A16

 

 

 

 

 

 

0856

A17

 

 

 

 

 

 

0744

A18

 

 

 

 

 

 

0178

A19

 

 

 

 

 

 

0200

C.1m

6/27/12

ok

ok

ok

-0.04

1.0192

0197

C.2m

7/3/12

ok

ok

ok

0.74

0.9900

 

M.05m

 

s1

 

 

 

 

 

M.1m

 

s5

 

 

 

 

 

M.2m

 

s9

 

 

 

 

 

M.3m

 

s12

 

 

 

 

 

M.4m

 

s14

 

 

 

 

 

M.5m

 

s16

 

 

 

 

 

M.10m

 

s7

 

 

 

 

 

M.20m

 

s10

 

 

 

 