Nov 30, 2010.  Visited several sites to install recalibrated kh2o's.

Site 2 (ABC) 1389 head bad; output ~30 mv.  Installed 1393 about 10:30; output ~2 V.

Site 3 (Highland) 1333 head  bad; output ~30 mv

                          1258 head bad; output ~20 mv

                          1390 head bad; output ~20 mv

                          1525 good!; output 2.2 V

Attached 1133 head to 1525 electronics and warmed head; still bad.

Reattached 1525 head; output 2.3 V.

Reattached 1525 head and left.

Site 1 (Playa) Installed 1395.  Also tested 1394, which was good.

Site 7 (Riverton) Installed 1394 around 15:55; output ~2.3 V.

Dec 1, 2010:  Completed kh2o installation.

Cleaned 'bad' kh2o's by scrubbing with qtips and distilled water. 

1133 went from 30 mv to 2 V

1258 went from 14 mv to 2 V

1389 went from 23 mv to 2 V

1390 went from 11 mv to 1.7 V.  Designated as spare.

Site 4 (West Valley): 12:10 installed kh2o 1258; 2 V output

Site 6 (West Slope): 14:30 installed kh2o 1389; 2 V output

Site 5 (East Slope): 15:50 installed kh2o 1133; 2 V output