Steve found an error in the NR01 coefficients that affect radiation measurements from SAVANT. As part of the process of correcting the measured values, we needed to determine the serial numbers for all NR01's deployed and track any swaps that may have been done in the field.

 

As it turns out, no NR01's were replaced during SAVANT, based on serial numbers reported in the raw data. The table below includes the serial numbers recorded in these data, which have also been added to the sensor list.

SiteNR01 Serial #Changes?Requires Coeff. Correction?
Rel - corn24NoYes
Rel - 20m26NoYes
Uconv - grass25NoYes
Lconv - soybeans29NoYes
Lconv - 20m19NoYes