This morning’s data check found no issues, other than the intermittent data from S11 (with poor cell reception).
Finish the soil for S16, and realized that we forgot to measure the boom angle for S16.  The boom angle will have to be measured during the next  service visit, or the last soil sample.
Unfortunately, the EC150 for S16 is still reading maxed CO2 - after the replacement of the scrubber bottles.

At Base, finished inputting calibration coefficients into TRHs.  We now have 7 spare TRH probes, all are in the truck console.
Looked over TRH bodies, and tried to determine what was wrong with the spare unit that was in the truck (ended up not having 12V on the wires to the Fan). Moved this unit to the bad pile.  We now only have 2 spare TRH bodies, both are in the truck with spares.
Tony has been fixing some of the TRH bodies.  If we have anymore TRH body failures in the next few days, Tony will fedex a few to us.  Otherwise he will pass a few TRH bodies to Jacquie for her to bring out as spares.