On Feb 4, logged into system from Boulder to verify the serial numbers of the licors:

port

SN

assumed height

2

1163

2m (was at 30m

7

1166

7m

20 (was 17)

1167

16m

14

now disconnected

 

11

1164

43m

This agrees with the configuration previous to Feb 2, and the changes of Feb 2: move of the 30m unit (SN1163) to 2m, and the move of the cable from the 16m unit (SN1167) from port 17 to port 20 on the new data system box.

To check the licor serial numbers do:

adn # shut down data process
minicom ttySN
ctrl-A F # sends break
(Outputs (RS232 (EOL "0A0D"))) # enable carriage return in output
(Coef ?) # display serial number
(Outputs (RS232 (EOL "0A"))) # new-line termination, resume output
ctrl-A Q # quit minicom
aup # restart data process
ds
ctrl-C # verify sampling rates