Gordon, Nov 19 18:30 MST

At 22:39 UTC (15:39 MST) this TRH started generating the hex gibberish:

2010 11 19 22:35:30.1623  0.9771      29 TRH15 14.10 23.92 5413 732\r\n
2010 11 19 22:38:07.1683     157      29 TRH15 14.07 24.13 5410 738\r\n
2010 11 19 22:38:08.1425  0.9742      29 TRH15 14.05 24.19 5408 740\r\n
2010 11 19 22:39:53.3761   105.2      47 TRH\xff\xff1\xff\xff5 \xff1\xff4.\xff\xff\xff03\xff 24.\xff4\xff8 \xff5\xff40\xff\xff6 74\xff\xff8\r\n
2010 11 19 22:47:55.4524   482.1      42 TRH\xff\xff15 13\xff\xff.99 \xff2\xff4.26\xff \xff540\xff\xff\xff2 74\xff2\xff\r\n

I noticed the gravestones in cockpit after I had installed the new nidas code.

Powering it down and up with eio got it working again at 00:39 UTC Nov 20 (17:39 MST)

eio 5 0
eio 5 1

Perhaps we should do the power cycle of the TRH (port 5) every so often (10 minutes) at this site?

Kurt and Ling were at this site today, Kurt was working on the power inside the box, which had the low voltage problem.

I'll wait to see if the problem returns before doing the watchdog power reset on the TRH.