You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 2 Next »

15 m Li-COR (sn1176) dropped out of service on November 6, 2011.  The diagnostic value dropped to 215 indicating that Bit 5 signaled that the instrument's chopper motor failed. There was an indication a number of hours earlier that the chopper motor was beginning to fail, see: 

http://www.eol.ucar.edu/isf/projects/BEACHON_SRM/isfs/qcdata/plots/2011/11/06/licor_20111106.png

Chris and I made a trip to MFO to move the 7m Li-COR (sn 0813) up to 15m and to remove the dead instrument from 15m.  The tower is therefore operating again with three levels of water vapor and CO2 fluxes (2m, 16m, 43m).  See Chris' entry for details of this instrument swap and timing.

  • No labels