RF03

5/24/25

Tech: Ivana


Pre-flight

-Removed right dewpoint head, purged with air (visible moisture cleared out), tightened screws and fittings (1 out of 3 hex screws were loose; fitting from pressure transducer to head was tightened 1/4 turn)

-Removed left dewpoint flag, purged with air, confirmed dewpoint head screws were tight

-Checked radome lines for moisture


Flight

-CAVP_DPL and CAVP_DPR within 10-20 hPa (issue from RF02 resolved)

-Data system operating properly and saving to correct disk (issue from RF02 resolved)

-0712: ACD_FO3 seems to have stopped recording, and the value is red on Controls/Status page; Teresa walked Steven through steps via HexChat to get it running again


Post-flight

-2DS Humidity: 10.6%

-Used new thumb drive in attempt to save tech notes from ADS station; unable to use or eject disk due to configuration issue; required admin authentication; still had to force power off as monitor got stuck in shutdown mode