Microsoft Excel Spreadsheet PostTRHcalinfo.xlsx Aug 30, 2013 by Steven Semmer

The TRH sensors were run through the calibration lab after SCP. Temperatures were checked over the range of -20C to 30C. RH was checked over the range of 10% to 90% at a temperature of 7C. The table below presents the mean error, Sensor - Ref, and the minimum and maximum error.

For humidity, in the cases where the range of the error is large, the errors typically increase systematically from the smallest errors around 10% RH to the largest errors around 90% RH.