March 14, 2011
TESTED GOOD: TS016, TS017, TS015, TS013(cleaned connector), TS005, TS011.
BAD: TS002, TS004, TS003.
TS012 tested good but had #4 sensor out.
The kryptons were checked in the chamber at 10C. The density range went from 1 to 9 g/m^3.
sensor |
error range (g/m^3) |
---|---|
1525 |
1.52 to 3.00 |
1389 |
1.01 to 1.56 |
1390 |
0.65 to 0.81 |
1393 |
-0.21 to 0.02 |
1394 |
2.49 to 3.43 |
1133 |
1.88 to 2.47 |
1258 |
1.95 to 2.32 |