qctables shows that lately (not last week), A4 0.5m TRH was the highest RH.  Replaced SHT.  A10 0.5m TRH was the lowest RH.  Replaced SHT.

Also, Gordon noticed that A17 0.5mTRH was coming in as S/N87, with id=0x11.  This SHT was labeled 068!  Also, Gordon's serial number logbook entry shows a different S/N 2 weeks ago.  Using mote_dump, we don't see a 0x10 sensor, however statsproc has been creating NetCDF files with data for both 0.5m and 2m TRHs.  This is all very odd.  Replaced this SHT with 060 and data now coming in ok.