Another day of mostly messing with the Picarro:

- Swapped in a new ADM242AN RS232 interface chip into the Picarro's rotary valve controller.  This got it working!  Reinstalled and the Picarro now is able to run a inlet sequence.

- Set final inlet locations and secured them all.  Checked that fittings were tight and that filter packs had the paper filter insert.  Did timing tests using Picarro's clock (via walkie-talkie) by breathing into the inlet.  Since we were running at 2.5 lpm, the flow rate was about 0.5 m/s -- I walk faster!

- Also blocked the sonic at 0.5m.M in an attempt to verify that the Picarro clock is synchronized with our DSMs.  Started blocking with my hand at 02:38:36 and held it there for a bit longer than 10s.

- Somehow in the process, the Picarro stopped sending data out the serial port.  I've restarted (adn/aup) and verified that we are still prompting for the data.

Also about 03:10 started a cleaning sequence: kh2o.5m, li7500.2m, li7500.1m, radiometers.  Didn't bother with kh2o.10m, since voltages have been good.