January 9, 2013

Following the SCP project, we tested each sonic in a zero-wind enclosure within the EOL temperature chamber, measuring wind component offsets and sonic temperature errors over the nominal range -30 C to 55 C.  The following table shows wind component offsets as okay if the u and v offsets are less than +/- 4 cm/s and the w offsets less than +/- 2 cm/s (between - 20 C and 30 C).  The sonic temperature corrections are listed as the slope and offset of a linear fit (between -20 C and 30 C), tc.sonic = offset + slope*Tc, where Tc is calculated from temperature, relative humidity, and pressure measured in the zero-wind enclosure.

Sonic

Location

Cal date

u

v

w

tc offset (degC)

tc slope

0200

c.1m

6/27/12

ok

ok

ok

0.04

0.9812

0197

c.2m

7/3/12

ok

ok

ok

-0.74

1.0101

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 