You are viewing an old version of this page. View the current version.

Compare with Current View Page History

Version 1 Next »

We have found a bug in the humitter code where the output format is incorrect for temperatures between 0 and -1 C.  This causes both T and RH values to be reported as missing.  Examples are:

2010 11 12 12:28:44.0730  0.9708      29 TRH20 0.00 66.96 4004 2164\r\n
2010 11 12 12:28:45.0522  0.9793      30 TRH20- 0.02 66.95 4001 2164\r\n

2010 11 12 13:53:16.4885    0.97      30 TRH20- 0.99 74.28 3906 2430\r\n
2010 11 12 13:53:17.4685    0.98      30 TRH20 -1.01 74.27 3904 2430\r\n

The humitters thus will have to be reprogrammed.  Current thinking is to send all of them back to Boulder for Steve to fix.

  • No labels