Post cal results of temperature, temperature range 15C to -20C

Sensor

mean (error)

sd (error)   

min to max

017

.019

.014

-.036 to .001

023

.017

.012

-.004 to .030

019

.016

.014

-.001 to .032

013

.000

.006

-.007 to .010

025

-.008

.011

-.034 to .002

024

.002

.008

-.011 to .013

009

-.035

.007

-.044 to -.022

028

-.018

.007

-.028 to -.004

027

-.011

.016

-.027 to .02

Post cal of RH, two runs from 10% to 90% at 10C and -4C, Gortex still on

Sensor

mean (error)

sd (Error)

min to max

017

 

 

 

023

-1.7

1.2

-3.8 to -0.4

019

 

 

 

013

0.1

0.5

-0.7 to 0.8

025

0.7

1.44

-2.2 to 2.2

024

1.1

1.7

-1.1 to 3.8

009

-0.6

0.6

-1.1 to 0.9

028

0.6

1.9

-3.6 to 2.6

027

1.2

1.6

-2.1 to 2.6