10m.c (SN 1473) read high and 15m.c (SN 1390) read low during the project, so we want to know what is going on.  Some input:

  • The edges of the lens on 1390 were cloudy when inspected in the lab
  • In the staging lab, I fired up both along with SN 1494 which had been in the lab.  The h2o values were:
    • 1494 8.68 g/m^3
    • 1390 3.60 g/m^3 (5.08 lower)
    • 1473 9.21 g/m^3 (0.53 higher; 5.58 higher than 1390)
  • We just did the lab zero/span test.  During that:
    • 1473: zero 0.71 (3.33 higher than 1390); then zeroed
    • 1390: zero -2.62; then zeroed
    • h2o span is still in progress
  • After the zero/span, I'll repeat the staging lab test to see if anything has changed!