14 Jan: Militzer & Oncley

Purpose: fix rad.mote which died at 8:30 this morning (with Vmote = 12.5V)

(Huge funeral going on)

1415: Mote blinking normally; voltage okay; correct ?ID; not being receieved.  Cycled power, which reset Xbee and all now okay.  Local data store probably is okay. Would base mote reset (initiate with "eio 6 0/1") have solved this problem?

1420: Took NESW photos

1424: wipe off radiometers (just a few drops on uplooking domes); .outs dessicant looked good -- didn't check .ins. (just replaced 2-3 days ago)

kv = 2.0, so just fine.

TRH fan seems happy.