Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

We have done the following:

Wiki Markup
1> cleaned all radiometer domes.
 
  Decided simply to breath on the domes and wipe off the condensation.
 
  Order was Rsw.in Rlw.in Rsw.out Rlw.out.
 
  The sun was shining during this period, so the effect of cleaning should be measurable
.  All domes looked pretty clean before and after. 
.  ?\[PS: have looked at the data.  Haze made the radiation somewhat variable, so my best estimate is that cleaning changed Rsw.in by 2--10 W/m2.  Not very much.\]  All domes looked pretty clean before and after.  (2:20pm)

2> As per the task list, disinterred (at a cemetary!) the single Tsoil to check/correct its orientation.  Found that it was nearly horizontal at about 2cm depth.  Reinstalled it at an angle that I eyeballed to put the sensor from 1--6cm depth.  Before doing this it read 0.3 C, after 0.81 C.  Note that these temperatures are slightly above freezing.  I noted a thin (~1cm) crust of frozen soil, but underneath definitely wasn't, which all is consistent.  I should also note that there was about 10cm of snow covering the whole area that I attempted to replace when I finished.  Finally, I note that we first unplugged the TP01 while determining which wire was which.  Thus, its values reset to 0, but later appeared to be okay.  (2:32PM)

...